Hello, and welcome to TX Chats, a podcast program presented by the American College of Toxicology, also referred to as ACT. The mission of ACT is to educate, lead, and serve professionals in toxicology and related disciplines by promoting the exchange of information and perspectives on applied toxicology and safety assessment. These podcasts are part of an ongoing effort intended to cover new developments in the field of
toxicology and provide a platform for discussion of topics. Toxicologists, regulatory scientists, and others interested in the field may benefit from. The views and opinions expressed in this podcast are those of the guest speakers and do not necessarily reflect though official policy or position of the American College of Toxicology. We'd like to thank you for listening and are happy to present today's featured podcasts regarding the safety evaluation
of medical devices. I'm your host, doctor Mary Saunders, and joining me as co host is doctor Adelude Frempong. Today we are interviewing doctor Shane Gadd and Samantha Gadd of GAD Consulting Services. GAD Consulting Services is an internationally recognized toxicology, drug and medical device development consulting practice, serving primarily the pharmaceutical and
medical device industries and the areas of preclinical and early clinical evaluation, safety assessment, and risk assessment. The practice is primarily focused on technical and regulatory support of new product development and the technical support of existing products. Doctor God is the principle of God Consulting Services and has more than forty six years of broad
based experience in toxicology, drug and device development, statistics, and risk assessment. He has specific expertise in neurotoxicology, in read through methods, inhalation toxicology, immunotoxicology, and genotoxicology. A past president of the American College of Toxicology and the Roundtable of Toxicology Consultants, and the recipient of the two thousand and
eight ACT Lifetime Achievement Award, Doctor Gad has authored or edited fifty three books, eight independent chapters, and more than three hundred and fifty papers and abstracts in the above fields. He has contributed to and has meaningful personal experience with i INDs one hundred and twenty four successfully filed to date, NDA BLA A and DA five one ten k ID CCE, CITD, dietary supplement and PMA
preparation. Doctor Gag is fully experienced with the design, conduct, analysis, and reporting of preclinical and clinical safety and pharmacoclinetic studies for drugs, devices, combination products, and dietary supplements, and with regulatory submissions associated with them. Samantha is the Signior Director of Biocompatibility at GAG Consulting Services, where she has
supported clients with risk assessments of both drugs and medical devices for the last thirteen years. She holds a degree in chemical engineering from the University of Cincinnati, a background which led to her strong interest in the use of cusile methods and material chemical characterization for the evaluation of biological risk. Some has broad knowledge of leachable and exstructible testing for both pharmaceuticals and medical devices, as well as gas
emission testing for reading gas partway devices from study design through a risk assessment. She frequently consults with companies to ensure device by compatibility evaluation, testing, and risk assessment methods meet current regulatory expectations. During the first part of this episode, we will disc tell us the diverse medical device categories for safety assessments. In a bit of approaches conducting safety studies and differences in criteria used by regulatorcal
authorities in the US, Europe, Asia, Africa and Australia. During the second part of the episode, we will discuss transitioning from pharma to medical devices as a future carrier path in toxicology. Welcome to the podcast, Doctor Shane Gadd and Samantha Gadd. Thank you for being here today. Our first question goes to Samantha, what are the medical device categories safety evaluations? The safety
evaluation of medical devices is really driven by the level of control that's going to be needed to assure safety and effectiveness. For example, in the US, devices can be Class one, which has the least regulatory control, Class two, which requires more special controls, and class classes which are subject to the most regulatory controls. The device class will also determine the type of submission or
application that's going to be required to receive clearance, and that also guides what should be done in the safety evaluation. So most Class one in Class two devices or most Class one in some Class two devices can be exempt from pro market notification or fact in case submission not necessary to notify of the device on market or changes but this exemption is still subject to some limitations. You can
determine your device class way going to FDA's website or sorry the CFR Class two two devices or subclass one that aren't exempt will require five to ten K pre market notification and then Clus three devices, which require the greatest level of control. More invasive devices require propercal market approval application, which requires more testing occasionally
the investigational device evaluation where you're testing your trials with human subjects. So it does depend on the class of device. Most companies go a five to ten K route for approval, which requires ensuring substantial equivalent. Substantial equivalents means a device is as safe and effective as the predicate. So this is why I
started with the different classes. Advices is showing substantial equivalents for a device. This is a very common route that companies take. Means the new device is as safe and effective as a predicate. It does not mean the device the new and the predicate need to be the same, but they do have to say to saving use and then any differences in technological characteristics or other characteristics can't
raise any different questions of safety. So when you have a device and you wish to use the fact that can pathway and claim there's a predicate device, how do you show it to safe and effective? And that's really where then probably your question is meaning to go and I sidetracked, But that's where your I said in my nine three standards come in which international and also eighteen five sixty two. There are several device type specific guidelines out there that some devices
require very little evaluation, other others require a lot more, and the usually can be found in the ISO standards. Doctor gad, what are the typical endpoints evaluated? Well, there are four major categories or groups of safety evaluation testing non clinical that is, not in people, in vitro and in vivo
bioassay endpoints. These are largely derived from test sets that were first proposed and adopted in the nineteen sixties. It's part of the US from COUPIA guidelines and these studies are done primarily in fact across the board setting done. Its primarily you're done not unnecessarily on the devices themselves, but rather on extract the devices. The concept being that uh, the interaction biological interaction with communication and device
really is that the surface or materials which are extracted or drawn out of the device and have subsequent systemic contact. Number two would be the interactions of implanted devices with surrounding tissues and the bodies immune systems, the innate and adaptive immune systems, and these can range from devices which have contact with unbroken skin to devices which are implanted in critical organs in the body. The third group are
modifications of traditional toxicology endpoints, and again these primarily use extra of the boat the device itself. There are some forms, generally now not favored by the regulatory bodies, where a device can have direct contact with the test system, be it a bacterial culture, root, cl culture, whatever. And then number four chemical analyses that determine and quantify extractable materials and compounds that come from
the medical device, followed by usually liiature based risk assessments of the detected compounds, though recently there's also the use of computer modeling quantitative structure a typically relationship modeling, and there is no no real data on the device per se. Now, across all of these you need to consider a couple of precepts, if you will. The ten nine nine three testing that's done the ISO standard
ten nine nine three testing. The degree of testing, the nature is dictated by the nature and the duration of patient contact. Okay, Number two, we don't really have GOP qualified test materials here. Medical devices they are tested are GMP good manufacturing practice. And that reflects the fact that for devices where
unlike pharmaceuticals or pesticides as such, there isn't a focus on precisely identifying chemical moieties beforehand or this device has these chemical characteristics, and that leads to the next precept, and the precept is what's tested for medical devices is the final form medical device that's going to have patient contact, Okay, right up,
packaged or taken out of the packaging, it has sterilized. If it is to be sterilized by the same means that the device which will go on humans is. And one other big difference is whereas in UH in the region of where the regulation of pharmaceuticals medical devices, the submissions are not yet required to
be electronic. This is coming, but send and electronic submissions to data sets isn't something that has common to medical devices and across the board that is really what you what's required for testing medical devices. The guidelines are set that whether they are ISO ten nine ninety three or ISO eighteen five six two. There
are a couple of others that are part of the isosystem. They are guidelines, but they tend to be somewhat more rigid in their application. Then it's the case necessarily with drugs okay, and they are This fits back to remember the first thing I said is the first place where device testing came from were USP guidelines that just like the USP they BIOASSEA guidelines are really set to where
you do this, and it has a criteria that if you pass it, if it meets that criteria, passes, and if you don't, it fails okay, And there's not a lot of detailed interpretation within that. The second part or the next part to consider is that, uh, the aggregate of all the things you see in the device test set, the data that supports the device being safe for use for humans patients in the manner in which it's
intended to be used, okay, is dictated by these guidelines. But when you get to the end, it's considered as a complete set almost all devices you're going to do Side of toxicity testing on side of toxicity was in the early sixties brought to use first for in the USP for containments or closures for
draggings as we went from the glass bottles plastic and UH CONDATD for those that we have transferred from a container. OUH. And side of toxicity was was meant to be a look here, you can fail side of toxicity and your device may be quite approvable and quite safe because it's one tool and it's the exact application. Uh, this is not is not such as you don't consider
other data that you have. Okay, So it's a complete data sets that is utilized. Uh. That really is Uh is in a in a short stroke, Uh, what's involved in device testing? Uh? It is typically shorter in regiment, it's typically less expensive, but the product lifespan for medical devices also short. Had devices typically are loves expensive. So UH that completes
uh consideration of what's involved the testing for medical advices. Thank you? So some if I can ask you the next question, man, have there been any novel or innovative approaches to safety evaluation that is now accepted by the FDA? Well, really, the whole device file compatibility evaluation idea is relatively new and it's definitely evolving quickly and think regulatory agencies and industry and labs. I
would get meant the device manufacturers as consultants, but we're all kind of doing this together. There are lots of there are lots of call between regulatory agencies. There's lots of collaboration, which is something I really love, so I would I would say having the framework which guidelines and standards have provided have been huge. They have also capped away for the use of chemical characterization or quantitative
structure activity relationships you SAR and in vitro to maximize use of those alternate methods of evaluating. For example, chemical characterization can take the place of some systemic toxicity endpoints that can significantly reduce the number of animals that would be necessary. It's a really interesting field. I would say for FDA specifically, not what
do they now accept, but I will give the advice to check recognize and censors recognize the Census Standards database that is cd H FDA relevant. They have certain things in the ISO guidelines that they do or do recognize or in other DAD documents. If I use it most for the ISO series of guidelines, that you can look it up and then I'll tell you what is recognized and
what is excluded, and that can be pretty important when you are designing a study or evaluating the data. Yes, I'm really I really enjoy this stuff. I'm excited to be involved. Right now. All the industry is all kind of learning from each other and growing, Doctor gad. How does safety assessment differ for regulatories in the US versus Europe, Asia, Africa and Australia.
Okay, A good question. The first thing directly to grasp is that the ISO ten nine ninety three guidelines are almost universally. They're accepted by all the regulatory units that you or entities that you introduced. They all follow ISO. But the devils and the details they are some exceptions, some areas where
they have for whatever reason, chosen to you can a guideline will say you can use this as in this test, and yet some countries that's not so. The one example that would would stand out to me right now is sensitization Japan, the US, UH, China, Australia, all Europe are except
for Europe all wants tee. The Guinea Pig maximization study having the alternatives pretty well uh not accepted in Europe. There are a set of three and virtual alternatives that are an aggregate accepted for sensitization. That's probably the biggest single difference between them. Their minor nuances in some of the regions. Other than that
in terms of the test to be done, however, the perhaps the biggest differences that whereas in the US and Japan, typically they observed grandfathering of test results. If you test the device according to the standards, and it's like guidelines that are opper at that point that device has approved and unless there's some brilliant new science or something that says we missed something major, that approval will
stand through time. That's not the case in Europe. In Europe and including the UK, there is periodic review usually every three years of test sets they've done for a device. And if the guidelines for testing that says you will do this and this change significantly or in the mind of the reviewer, eyes of the reviewer changes, they then the testing has to be done that to
meet that new guideline. That's a significant difference that re examinations through years. Also testing submissions in the US and UH in China it's a little bit more complex there and elsewhere are to a government regulatory authority the FDA, United States, UH, Canada, Health in Canada, and so forth. UH.
In Europe, however, medical device testing UH and acceptance is reviewed by what they're called notified bodies, which are really private non governmental organizations UH and those will review and aggregate UH the submission UH and will approve we're not okay, and that approval by one of the regulator by one of these five bodies. Poles crossed Europe, Okay, they currently UH the European submissions are a little
bit in the state of flux because the Europeans have come up with what they call the mb R regulation, which are changing things and perhaps being more rigorous and some of the expectations. The other part is that in Europe the it's an opinion rendered, it's part of the submission by an expert by his or her qualifications are included this submission. That's not the case in the US, or Japan, India or such Okay, well all, however, all of
these require that testing be done in conformance with the guidelines, okay, and the guidelines that are enforced. Remember is the initial approval the guidelines that are enforced at the time the device is presented and evaluated and be done in conformance with good laboratory practice JP and pedetnations. There is a little nuance here also though that China is not a member European jlps are largely based around OECD.
China is not a member of OCD, so very recently European nations are saying they're not going to accept testing done in China that is not concurred in here. The compliance of testing, I think that covers all the parts the other nations that are there in Africa, and I did mention India is in confartment. It's also India and China, but Korea and Latin America. Again it's
typically the isoicide lines which which govern everything. One exception here, by the way, is that so ten nine ninety three TECH one table. Remember I said that testing was in conformance with the type of exposure, patient HATH and the duration. But if you look at that table, there's no coverage of respiratory devices. And of course with COVID we had a lot of concern with respiratory devices. That's rather governed by eighteen five six two. I saw eighteen
five six two, and it's subparts and to some degree by some APA and other regulations in terms of levels of matels that there may be exposure to. I think that pretty well covers the subject. Thank you. During the second part of this episode, we will discuss transitioning from farm to medical devices as a future career path in toxicology. There is tremendous interest among ATT colleagues on
transitioning from farmer to medical devices as a future career path in toxicology. What advice can you share to help scientists and will to pursue this career path. I'd like to start by offering perhaps a different perspective than other people, and that is I started in toxicality almost fifty years ago, when the majority of toxicology testing was for agricultural chemicals and occupationals chemicals and exposures, perhaps also something
environmental. My good fortune and not great great common sense managed to make transition in the late seventies very early eighties from that field to pharmaceuticals, and later on I had the alternative, the opportunity to also become involved in the early stage with safety evaluation for medical devices. The lesson there is there is no
final place you're going to end up. You have to You have to be able to change as societies requirements and regulatory requirements move to a different area. Used to be all the contract research organizations are out there depended really on chemical and Agrica chemical industry and agricultural testing or evaluation of that's no longer the case
right now. The maturity are for therapeutics, pharmaceuticals, biochemical or a biotechnology products and so forth. But keep in mind that the medical device field the market is about the third the size that drugs. It's not small, and there's a lot of opportunities there. I have said, I'm going to turn it over to Samantha. Yeah, I also have a couple two cents to
share for that one. So what advice can I share help people were interested in pursuing getting more into medical devices. I'd say, opportunity isn't going to come to you. If it does, you should be ready for it, So start learning and reading. I spend way too much time reading articles and looking at guidelines and presentations on the internet webinars. There are a lot of
device industry relevant meetings as well where you could you go make yourself seen. But as Actor has said, many many device companies and for many device companies, they have devices on the market which weren't we're hard to pass or go through the testing that we have in place currently as requirements. So many of those companies are are you know, kind of wanting to UPDI stuff and going, oh, hold on, what do you mean I have to do you
what do you mean I have to do testing? So there's a lot of there's a lot of growth in the device industry. And I just know what you're talking about when you get in there, because I lot of little nuances if you get confusing between drugs and devices for some things, but if you really try to work your head around where where the industry is right now and how do we evaluate a medical device and how it interacts the body? And Yeah, also you know ACT has even been bringing a lot more medical device
topics to the meetings. There's going to be a continuing education at this year's annual a meeting on medical devices that should actually be very very interesting. We've got a good group of speakers including industry and the FDA, and it's going to be a really good talk. Just get out there and see what's available.
But read constantly read, it's constantly changing. You have to keep learning, and you have to read that what you know today may not be what the field expects or what the regulator expect. What What are the transferable skills toxicologists need to be successful in medical devices? Basic understanding of chemistry and organism, level biology, human physiology, and podcast mystery. A ability to critically
evaluate data and to communicate with those people that need to understand it, either your senior management or a potential customer or regulators. Okay, those are probably the fundamental skills that are required. And recognizing that change is the eternal verity.
You're going to have to be able to change and understand the world in which our base of knowledge continues to grow. Medical devices, when I first got into them the very early eighties, were pretty fundamental, and people still when you say while I work on medical devices, they think bandages and perhaps
stints and such. But now we have medical devices which monitor levels of infection or levels of vital biochemical components such as insulin in the body and not only measure those evaluate them, but may also disperse therapeutic agents. Okay, so the technology continues to change, you have to be able to at least understand
the change technology the time comes when you can't be an expert in everything. In fact, the time we've properly start. But you have to be able to understand and be able to communicate with those people whose expertise is fundamental or critical to what you're doing. Well. To answer the question about transferable skills of a toxicologist, I guess I have to take a different approach because I
came into all of this as a chemical engineer working in a toxicology consulting company. So my my transferable skills I needed for medical devices where really similar my doctor katsaid and just flat of mental chemistry. For me with the background I have and what has made me so comfortable comfortable, but when it just made me really enjoy medical device evaluation has been more so the draw from my background.
And I know that that's somewhat of a common thing we're seeing bio compatibility engineers or whatever engineers, and but it really has been helpful. Helps me to understand the material properties and the results of the chemical characterization. I mean, that's still part of the evaluation. Your chemical characterization is replacing a lot of stuff, and you need to be aware of what you're doing and what the data represents and how to evaluate it. So really, I think for
me, the the best most important part has been honestly having a lab walk me through the report and extractable mutual report to explain it to me once I'm like eleven years ago and then I was like, oh, okay, that's what they're doing. It's you know. Being able to understand the knowledge and the the correlation between DO testing and the material with the human body has been
extremely helpful. Still have sat Anse. Thank you for this opportunity to participate in today's podcast. Thank you so much for both of you. We really appreciate your time and your contributions, and we look forward to seeing you at the ACT annual meeting. This podcast was brought to you by the American College of Toxicology, whose mission is to educate, lead, and serve professionals in
toxicology and related disciplines by promoting the exchange of information and perspectives on applied toxicology and safety assessment. For more information, please visit www dot actox dot org.
In this podcast, we sit down with Dr. Yvonne Maldonado of Stanford University School of Medicine, a widely recognized leader in infectious diseases. Building on the conversation sparked by Dr. Maldonado’s plenary talk,...
This episode of ToxChats is designed to offer listeners a practical and insightful overview of the waiver process for carcinogenicity studies in pharmaceutical development, with a focus on the ICH S1B and S1B(R1)...
In this episode of ToxChats, hosts Lauren Walker and Sarah Benjamin sit down with the groundbreaking Donald Ingber - visionary scientist, engineer, and Founding Director of the Wyss Institute. Dr. Ingber shares how his...
In this episode, Tanya McDonnell and Hemraj Dodiya sit down with two highly regarded toxicologists, Ric Stanulis and Mary Ellen Cosenza, who transitioned from industry and currently work as consultants, either...
In this episode, Aimee Altemus and Adelaide Frimpong speak with Dr. Gary Miller of Columbia University, a widely recognized leader in exposome research. Dr. Miller discusses how the exposome - the full spectrum of...