Let's Talk Risk! Podcast

Let's Talk Risk! Podcast artwork

Deep Dive: When the Risk Baseline and the QMS Both Break Down

Sep 25, 202622 minAudio
ShareXLinkedInFacebookWhatsAppEmail

The baseline is completely broken and the iteration process is blind. FDA’s August 21, 2026 warning letter to Pensar Medical LLC illustrates a failure that goes well beyond incomplete documentation. The underlying risk-management foundation was weak. Risk-control verification remained incomplete for critical areas, and FDA reportedly found no design, process, or use-related FMEAs for the WoundPro product line. At the same time, key QMS processes were also failing. Key highlights covered in the audio: * Risk analysis was incomplete or missing. Critical risk controls remained unverified, and some products lacked the basic FMEAs required by the firm’s own procedures. * Design and software changes were poorly controlled. Multiple changes lacked adequate implementation records, including an engineering change order described simply as “clean up all code,” with no verification record showing what had actually changed. * Field corrections were not properly managed. Software rework performed in the field was not reported within the required timeframe. * CAPA was not functioning as a learning system. FDA found CAPA records with missing root-cause analysis, corrective actions, and effectiveness verification. In another case, a corrective action was never incorporated back into the risk analysis. * Post-market information was not being translated into effective action. A serious injury involving retained foam and surgical intervention was reported late and categorized as a malfunction rather than a serious injury. The central lesson is straightforward: A robust QMS needs both a reliable risk baseline and processes that continuously challenge, maintain, and improve that baseline. If the original risk analysis is weak, changes cannot be evaluated properly. If CAPA, complaint handling, change control, and post-market processes are weak, new information never strengthens the risk analysis. That is how individual weaknesses become a systemic quality failure.. Keywords: Pensar Medical, FDA Warning Letter, QMSR, ISO 13485, Risk Management, Risk Analysis, Design Change Control, CAPA, Medical Device Reporting, Quality Management System 🎧Listen to the Deep Dive for a closer look at how the Pensar warning letter shows the consequences when both risk management and the broader QMS fail to operate as a closed-loop system. Thanks for reading Let's Talk Risk!. If you liked this post, share with others. Note: The audio summary was prepared using Google NotebookLM, an AI-enabled research tool. Here are a few key resources used for this analysis: * FDA (2026). Warning Letter: Lion Street Medical, d/b/a Pensar Medical LLC (CDRH-MARCS-CMS 733015). This is a public episode. If you'd like to discuss this with other subscribers or get access to bonus episodes, visit naveenagarwalphd.substack.com/subscribe

Audio and video play straight from the show's own hosting. PodSubs stores and serves no media.

Transcript

From the show. Tap a time to play from there.

SPEAKER_01Imagine authorizing your engineering team to push an update to a life support device and the work order literally just says, clean up all code.

SPEAKER_00Oh, yeah. Terrifying.

SPEAKER_01Right. I mean, no record of what was actually changed. No testing. Just, you know, a blind software update pushed out into the wild.

SPEAKER_00Just crossing your fingers and hoping for the best.

SPEAKER_01Exactly. And today we are doing a deep dive into an FDA warning letter that shows exactly what happens when a medical device company treats safety like a rough draft.

SPEAKER_00Which is unfortunately more common than you'd think.

SPEAKER_01Yeah. And for you executives listening, here is the absolute strategic bottom line on quality management system risk governance. Compliance is not a regulatory tax on your business.

SPEAKER_00Right. It's not just a hoof to jump through.

SPEAKER_01It's routine operational systems design. I mean, it as is what gives you speed to market, predictable product performance, and ultimately it protects your long term enterprise value.

SPEAKER_00Yeah, because when a leadership team views compliance as this separate annoying hurdle rather than, you know, the foundation of their strategy, they literally lose control of their own product.

SPEAKER_01100%.

SPEAKER_00I mean, the regulations we're going to unpack today aren't arbitrary rules invented by bureaucrats. They're codified engineering principles.

SPEAKER_01So I want to dig right into this because our source material today is an FDA warning letter. It was issued on August 21st, 2026 to Lyon Street Medical doing business as Pensar Medical LLC.

SPEAKER_00Right. They manufacture negative pressure wound therapy systems.

SPEAKER_01Yeah. Devices like the Wound Pro, the Wound Pro Plus A, and the MicroDoc, which is also referred to as the PocketDoc.

SPEAKER_00Yeah.

SPEAKER_01But before we get into the failures, let's just establish what these devices actually do because you really need to understand the stakes here.

SPEAKER_00Well, negative pressure wound therapy, it's used for patients with really severe, often chronic wounds. So think diabetic ulcers, severe burns, or like large surgical incisions that just won't close properly.

SPEAKER_01Okay, so highly vulnerable patients.

SPEAKER_00Exactly. And the system involves placing a specialized foam dressing directly into the wound bed, sealing it with an airtight film and connecting it to a vacuum pump. Right. So the device applies continuous or intermittent negative pressure, a vacuum effect basically, to draw out fluid or exudate and infectious materials into a little canister.

SPEAKER_01So it physically pulls the edges of the wound together while removing the fluids that cause infection. And that promotes blood flow and tissue growth.

SPEAKER_00That is the goal.

SPEAKER_01Yeah.

SPEAKER_00And because it interacts directly with compromised tissue, the mechanical precision of that pump, the sterility of the packaging, and the reliability of the software controlling the vacuum pressure, it's all absolutely critical.

SPEAKER_01Because if it fails.

SPEAKER_00If the pressure drops, the wound just stops healing. But if the pressure spikes, you can actually damage that fragile new tissue.

SPEAKER_01Which brings us to the mission for this deep dive. We are going to explore how a failure to integrate risk management across the entire product lifecycle led to a complete cascade of operational failures for Pensar.

SPEAKER_00Right. Looking at this through the lens of the QMSR, the Quality Management System Regulation, and ISO 13485.

SPEAKER_01Yeah. Specifically, what happens when risk management isn't built into that initial manufacturing phase, how software changes become incredibly dangerous without a baseline, and why a broken KP system basically guarantees you'll repeat your mistakes.

SPEAKER_00Well, the best place to start is at the very beginning of a product's life cycle. What the industry calls product realization.

SPEAKER_01Which is when it goes from a drawing to a real thing, right?

SPEAKER_00Exactly. It's the phase where you transition a design on paper into a physical product you can actually manufacture and sell. And Clause 7.1 has a very clear rule here. You must document risk management during this realization phase.

SPEAKER_01Right. And according to the warning letter, the FDA inspector looked at Pensar's design failure mode and effects analysis, their DFMEA for the pocket dock device.

SPEAKER_00And what did they find?

SPEAKER_01Well, the document did identify distinct hazards, but when it came to the actual risk control verifications, you know, the proof that they mitigated those hazards.

SPEAKER_00Let me guess, they were missing.

SPEAKER_01Entries for software, sterilization, packaging, exudate management, they were all simply marked in progress.

SPEAKER_00Oh, wow. In progress.

SPEAKER_01Yeah.

SPEAKER_00I mean, leaving a risk control marked in progress means you've identified a way the device could potentially harm a patient, but you just haven't bothered to prove your proposed solution actually works.

SPEAKER_01I am genuinely trying to wrap my head around this. If I'm a CEO and I see in progress on my critical risk controls for something like sterilization, my first instinct is, okay, this product physically cannot leave the building.

SPEAKER_00Right. It shouldn't.

SPEAKER_01Am I misunderstanding how FDA clearance works here? Or did this company just totally bypass their own gatekeepers?

SPEAKER_00Well, it's actually a super common misconception about how the regulatory framework operates. Oh, really? Yeah. The FDA clears the initial design of a device based on a pre-market submission, assuming the data you provide is accurate. But they don't, like, sit in your facility checking every piece of paper before a box ships.

SPEAKER_01Oh, so they're relying on you?

SPEAKER_00Completely. They grant the clearance with the legal expectation that your internal quality management system will act as the gatekeeper. Pensar essentially dismantled their own gates.

SPEAKER_01You know what this reminds me of? From an executive standpoint, it's almost like building a massive cargo ship.

SPEAKER_00Okay.

SPEAKER_01And you launch it into the harbor while the hull's watertight doors are still just like a sketch on a notepad.

SPEAKER_00That is a perfect analogy.

SPEAKER_01You might float out of the harbor on a sunny day, but the second you hit rough water, you are taking on water and going down. I mean, how does an operation even function without a basic risk map?

SPEAKER_00Yeah. Yeah. But what is truly staggering is that for their Wound Pro product line, the FDA found absolutely zero FMEAs of any type.

SPEAKER_01Wait, wait, zero. No design FMEA, no process FMEA, no user FMEA.

SPEAKER_00Nothing was found to exist.

SPEAKER_01It's just nothing. Wow.

SPEAKER_00And their own formal QMS process risk management procedure required all three of those for every single product.

SPEAKER_01So they wrote a rule book for themselves and then entirely ignored it. I really want to dig into the how of this. Like, how does a company arrive at a point where an entire product line has zero risk documentation?

SPEAKER_00Yeah.

SPEAKER_01Is it just sheer incompetence or is there some systemic operational flaw at play here?

SPEAKER_00Well, in many organizations that struggle with this, it really comes down to a siloed culture.

SPEAKER_01Okay, how so?

SPEAKER_00So research and development might design a product, get the regulatory clearance, and then basically toss the design over the wall to manufacturing.

SPEAKER_01Ah, I've seen that.

SPEAKER_00Right. If R&D never finalized the risk verification and manufacturing is, you know, only incentivized by production quotas, the paperwork just sits in limbo forever.

SPEAKER_01And the warning letter also noted that when the FDA pointed this out, Pensar's short-term correction was inadequate because they lacked an interim risk mitigation plan.

SPEAKER_00Yes.

SPEAKER_01Let's translate that for the executives listening. What exactly is an interim risk mitigation plan?

SPEAKER_00So if you're caught manufacturing a device without verified risk controls, you can't just promise the FDA, hey, we'll hire a consultant to write up the FMEA over the next six months while you just keep shipping the product as usual.

SPEAKER_01Because you're still putting unverified devices on the market.

SPEAKER_00Exactly. During those six months, you are still exposing patients to unverified risks. An interim plan means putting immediate temporary guardrails in place.

SPEAKER_01Like what?

SPEAKER_00Like 100% manual inspection of every single unit. Or even halting production entirely until the foundational paperwork is actually fixed. And Pensar just didn't do this.

SPEAKER_01So if a company doesn't know what its baseline risks are, and they don't have a verified map of how their product functions safely, what happens when they inevitably need to change or update that product?

SPEAKER_00Well, that's where things are really dangerous.

SPEAKER_01Right. That dynamic brings us to the next massive failure in this morning letter. Which is Clause 7.3.9, the control of design and development changes. Because without a baseline, engineering changes just become blind gambles.

SPEAKER_00Totally. The standard requires any medical device manufacturer to adequately review, verify, validate, and approve design changes before they're implemented. You have to prove the change doesn't introduce a totally new hazard.

SPEAKER_01But because Pensar didn't have their baseline risk controls locked in, their change management process was basically chaos.

SPEAKER_00Absolute chaos.

SPEAKER_01Let's look at the software side of this. The FDGA found they implemented multiple software changes to the MicroDoc and WoonPro Plus devices without any batch records confirming the updates actually took place on the manufacturing line.

SPEAKER_00It's just wild.

SPEAKER_01For instance, they had an engineering change order, ECO 25039, for a WoonPro Plus firmware update, but zero proof it was installed.

SPEAKER_00And during the inspection, the company's president actually confirmed that no such batch record had been documented.

SPEAKER_01Unbelievable.

SPEAKER_00But the most glaring example of this chaos was ECO 25008. The stated purpose of this engineering change order was to, and I quote, clean up all code.

SPEAKER_01Okay, I have to push back on this for a second. Because in the broader tech world, right, software engineers push massive code cleanups, refactoring agile updates all the time. It's standard practice to improve user experience.

SPEAKER_00Sure, in consumer tech.

SPEAKER_01Yeah. So why is a general code cleanup such a cardinal sin in medical devices?

SPEAKER_00Well, the difference really lies in the consequences of a failure. And just the interconnected nature of the software. I mean, in consumer tech, if a code cleanup introduces a bug, an app crashes and you have to restart your phone.

SPEAKER_01Right. Minor inconvenience.

SPEAKER_00But in a negative pressure wound therapy device, if a code cleanup inadvertently changes, say, the timing of the vacuum sensor.

SPEAKER_01Oh, man.

SPEAKER_00Yeah. The device might apply continuous high pressure to a fragile surgical wound. That can cause severe hemorrhaging.

SPEAKER_01And the warning letter states that neither the CEO nor the president could even explain what code changes were actually incorporated under that cleanup all code order.

SPEAKER_00No verification record existed at all.

SPEAKER_01Hearing an engineering order to clean up all code with zero paper trail that is terrifying for leadership, that is not a controlled change. That is introducing a black box into your product line.

SPEAKER_00Exactly. Which reveals a fundamental misunderstanding of regression testing within their software validation procedure.

SPEAKER_01Walk us through regression testing.

SPEAKER_00So regression testing is the process of confirming that unchanged portions of the software weren't inadvertently disrupted by a change you just made. Software is deeply interdependent, right? Right. So a change in a module that controls, let's say, the display screen could unexpectedly draw more power. And that causes the module that controls the pump motor to starve and shut down.

SPEAKER_01Oh, wow.

SPEAKER_00If you aren't performing regression testing against a baseline risk map, you are literally just crossing your fingers.

SPEAKER_01How does a team confidently drive product iteration and speed to market if they cannot prove what code is in the device they just shipped?

SPEAKER_00They can't. They also made a hardware or firmware change that they characterized as a cosmetic noise change without documenting a risk assessment tied to the original DFMEA.

SPEAKER_01They just said it was cosmetic.

SPEAKER_00They just declared it was unrelated to safety or effectiveness.

SPEAKER_01Let's break down the mechanics of why a quote-unquote cosmetic noise change is never just cosmetic in a vacuum pump.

SPEAKER_00Okay, so to make a mechanical pump quieter, you usually have to change the physical housing, perhaps adding dampening materials, or you change the voltage to lower the motor's RPMs.

SPEAKER_01And if you add dampening foam to the housing, you might trap heat, which could cause the motor to overheat and fail.

SPEAKER_00Precisely. Or if you lower the motor RPMs, you might change the flow rate of the exudate being pulled from the wound. You cannot simply declare a change safe without proving it against your risk map.

SPEAKER_01Every alteration has to be evaluated.

SPEAKER_00Every single one. You have to ensure it doesn't introduce a new hazard. Proper change governance isn't a roadblock designed to slow engineering down. It's the very mechanism that allows you to innovate safely and quickly.

SPEAKER_01And this engineering chaos, it inevitably bleeds into regulatory trouble. Specifically with corrections and removals, What happened with the MicroDoc software update out in the field?

SPEAKER_00So Pensar realized they had a virgin issue with the MicroDoc software from multiple customers. So they just conducted software reworks out in the field to get these customers on the correct revision.

SPEAKER_01Okay.

SPEAKER_00Now, under 21 CFR Part 806, fixing a software flaw in the field to reduce a risk to health is considered a medical device correction. And you are legally required to submit a report to the FDA within 10 working days.

SPEAKER_01Let me guess, they updated the software out in the wild without telling the agency.

SPEAKER_00They missed the 10-day window entirely. As of the date of the warning letter, they still hadn't submitted it.

SPEAKER_01Man! So we have unverified original designs combined with unverified undocumented code changes and unreported field corrections. It's a disaster. The baseline is completely broken and the iteration process is blind. In a healthy operational system, when things break in the field, internal feedback loops catch the issues before they spiral, right?

SPEAKER_00Usually, yes.

SPEAKER_01But this brings us to Clause 8.5.2 and MDRs. Basically, corrective and preventive actions. The final piece of this failure cascade. Because at Pensar, the feedback loops were completely broken, cutting off the final safety net.

SPEAKER_00Your CAPIA system is supposed to be the engine of continuous improvement for your business. When a failure occurs, you investigate it, you document the root cause, implement a fix, and then you actually verify that the fix worked.

SPEAKER_01It's basic problem solving. Yeah.

SPEAKER_00And the regulations explicitly require you to document these steps. But the FDA reviewed five of Pensar's Cappier records and found this severe pattern of blank fields.

SPEAKER_01I often think of CAPA as an enterprise's immune system. You know, when a pathogen, like a defect or a field failure, enters the body, the white blood cells attack it, learn from it, and create antibodies so it never harms the organization again.

SPEAKER_00I love that. But Pinsir's immune system was completely offline. CAPA 037, which was opened in January 2024, was entirely incomplete.

SPEAKER_01How incomplete?

SPEAKER_00The root cause section, the corrective action sections, and the effectiveness verification sections were all left completely blank.

SPEAKER_01How does that actually happen in a corporate environment? Is it software fatigue like quality engineers just opening tickets in a digital system to hit a geeky eye and then forgetting about them?

SPEAKER_00It's often a mix of template fatigue and honestly a culture of fear.

SPEAKER_01Oh, interesting.

SPEAKER_00Yeah. When compliance is viewed as just a paperwork exercise, an engineer might open a KPI just so they can tell management, hey, I opened a ticket.

SPEAKER_01Right.

SPEAKER_00Box check. Exactly. Another issue is that documenting a root cause feels like admitting liability to some people. If an organization doesn't have a culture of psychological safety, employees will literally leave fields blank rather than put in writing that the engineering team messed up.

SPEAKER_01Wow.

SPEAKER_00We saw this with KPAR Series 9, which contained only unfilled placeholder language that carried over from a template.

SPEAKER_01So they just opened a digital template, saved it, and never did the work. Seeing a completely blank KPAR is like your white blood cells clocking out for lunch during a severe infection. No, and that is where the entire QMS fractures. KPA 036 noted that a corrective action was taken. But that action was never incorporated back into the firm's risk analysis, and the effectiveness fields were left blank.

SPEAKER_01This is a critical point. If we want to build resilient organizations, how do you practically integrate risk analysis back into a KPA system so it actually works?

SPEAKER_00Well, it has to be a continuous closed loop system. When a failure occurs, your investigation must ask one vital question. Did we predict this failure in our original FMEA?

SPEAKER_01Okay.

SPEAKER_00If the answer is yes, then you have to ask why your predicted risk control failed. But if the answer is no, then your FMEA is incomplete and you have to update it with this newly discovered hazard.

SPEAKER_01But you can only trigger a KPAY if your internal definitions for recognizing a problem are actually clear in the first place. Which brings us to their medical device reporting, or MDR, failures. Right. Because an MDR is often the trigger for a major KPAY.

SPEAKER_00The regulation dictates that you have to report an event to the FDA if information reasonably suggests your device caused or contributed to a death or serious injury. But Pensar's internal procedure for MDRs didn't even include an adequate definition of what reasonably suggests means.

SPEAKER_01Let's look at the psychology of that, because reasonably suggests wasn't clearly defined. Imagine a quality engineer sitting in a cubicle gets a complaint about a wound infection. Without a strict definition, human nature takes over, right? They downplay it. They think infections happen all the time. The patient was already sick. It's just a device malfunction.

SPEAKER_00Yeah, that happens a lot.

SPEAKER_01They don't realize the regulatory standard requires them to assume the device contributed to the injury until they can definitively prove otherwise.

SPEAKER_00And that lack of definition led directly to the mishandling of complaint C-029. A patient using Pensar's negative pressure system sustained an infection at the wound site, specifically purulent drainage.

SPEAKER_01Which means it was actively weeping infectious fluid.

SPEAKER_00Exactly. The situation escalated to the point where the patient required medical and surgical intervention to remove retained foam from the wound to preclude permanent impairment.

SPEAKER_01Just so you listening understand the medical mechanics here, the foam dressing from the device was left inside the patient's wound. And when that happens, the body's natural granulation tissue actually starts growing into the porous structure of the foam.

SPEAKER_00It's awful.

SPEAKER_01You can't just pull it out like a Band-Aid. A surgeon has to go in with a scalpel and physically excise the foam and the new tissue.

SPEAKER_00It is a severe event. That is the very definition of a serious injury under the regulations. Now, Pensar became aware of this event on May 1st, 2024. They failed to submit the required MDR within the 30 calendar day window.

SPEAKER_01Typical.

SPEAKER_00And when they finally did submit it, they incorrectly categorized it as a malfunction report instead of a serious injury report.

SPEAKER_01How does a quality engineer downgrade a surgical excision to a mere malfunction?

SPEAKER_00Well, because the procedure didn't define reasonably suggests, they likely convinced themselves that the infection caused the surgery, not the retained foam, completely missing the fact that the device was inextricably linked to the event.

SPEAKER_01Which completely misrepresents the severity of the issue to the regulatory agencies and, frankly, to their own internal management teams looking at the data.

SPEAKER_00Right.

SPEAKER_01If the executive team sees a dashboard with malfunctions, they don't allocate resources. Right. If they see serious injuries requiring surgery, they stop the assembly line.

SPEAKER_00When you connect all these dots, you really see the breathtaking cascade of systemic failure. A serious injury occurs. It is miscategorized and delayed in reporting because the procedures lack basic definitions. Because the reporting is flawed, the severity isn't recognized by leadership. Therefore, the resulting KPAs are opened but left as blank templates or closed without effectiveness checks. It all stems from viewing risk management as an isolated theoretical exercise rather than the living, breathing operational system it is supposed to be. 100%.

SPEAKER_01Leaving KPAs blank or disconnected from risk analysis basically guarantees that you will repeat your failures. You will keep shipping devices with unverified software changes, and you will keep having retained foam incidents because you never closed the loop.

SPEAKER_00None of these FDA citations are about government overreach or nitpicking typos. They're about a company that lost total control of its engineering and quality feedback loops. A solid, interconnected quality management system is a strategic asset. When your risk controls are verified before launch, when your software changes are regression tested against the baseline, and when your corrective actions actually update your corporate knowledge.

SPEAKER_01Things just work.

SPEAKER_00Yes. Your business runs with incredible efficiency. You can innovate faster because you know exactly where the guardrails are.

SPEAKER_01That is the mindset shift that separates industry leaders from companies drowning in compliance issues. It is about maintaining absolute, verifiable control over your product's lifecycle to drive enterprise value.

SPEAKER_00And Pensar serves as a stark reminder of what happens when that operational control is entirely abandoned.

SPEAKER_01Oh, wow. Yeah. If an organization's quality management system cannot even govern static human written

SPEAKER_01code today, how in the world will it handle the autonomous engineering of tomorrow?

Open the transcript file

Hosts and Guests

Nobody is credited on this episode yet.

Credit the people on this episode

Be the first to rate this episode

More episodes

All episodes
  1. LTR 168: Leading Through Regulatory Uncertainty

    Nov 20, 202627 minAudio

    Summary “Be guided by your own values and facts. That’s important. That will keep you centered” Quality and regulatory leadership in MedTech is about much more than ensuring compliance. Senior leaders must navigate...

    Credit the people on this episode

  2. LTR 167: Making AI Work in MedTech QMS

    Nov 12, 202627 minAudio

    Summary “First you have to understand the technology, what the technology does, what it's capable of, and you have to understand its limitations.” Artificial intelligence is rapidly changing how medical device companies...

    Credit the people on this episode

  3. Executive Brief: Scaling Software as a Medical Device

    Nov 6, 202622 minAudio

    Software velocity is not simply a development challenge. It is becoming an enterprise capability. Software is increasingly driving differentiation, recurring revenue, and customer value in MedTech. But many...

    Credit the people on this episode

  4. LTR 166: Building a MedTech Career in the Age of AI

    Oct 30, 202626 minAudio

    Summary “At the end of the day, this business of getting a job, keeping a job, getting promoted is human to human.” The MedTech job market has changed. More candidates are competing for fewer well-matched opportunities,...

    Credit the people on this episode

  5. Executive Brief: PCCPs as a Competitive Strategy

    Oct 23, 202621 minAudio

    PCCPs can turn regulatory cycle time into a competitive advantage, but only if the evidence needed to support future change is funded in advance. Predetermined Change Control Plans (PCCP) are often discussed as a...

    Credit the people on this episode