Robust chemical characterization is essentially the gateway to biological safety. A toxicological risk assessment is only as strong as the analytical data supporting it. FDA’s September 2026 biocompatibility town hall provided a detailed look at how the Agency is evaluating chemical characterization and toxicological evidence under ISO 10993-18. The emphasis was not simply on whether testing was performed, but whether the resulting data are sufficiently representative, sensitive, and scientifically justified to support conclusions about biological safety. That expectation reaches well beyond testing. Extraction design, analytical thresholds, device variability, unidentified chemicals, surrogate standards, supplier changes, and CRO practices can all determine whether the toxicological evidence package remains defensible. Key highlights covered in the audio brief: * Characterize the final finished device. Raw-material data may not adequately represent the chemistry of the manufactured and sterilized device because processing can introduce degradation products and manufacturing residuals. * Use extraction conditions that are both conservative and clinically relevant. FDA’s examples show that an extraction can be too weak - but it can also become artificially aggressive and generate chemicals that do not reflect actual patient exposure. * Ensure the analytical method can actually support the toxicological threshold. FDA is scrutinizing how the analytical evaluation threshold, limit of quantification, and uncertainty factors are established and justified. If the method cannot reliably detect chemicals at the required safety threshold, the toxicological conclusion becomes difficult to defend. * Account for device-to-device variability. Analytical replicates measure instrument repeatability. Extraction replicates using separate final finished devices are needed to evaluate variability in the devices themselves. * Do not eliminate variability through pooling. Combining extracts can dilute a high-result outlier and mask an intermittent manufacturing-related chemical exposure. FDA specifically cautioned against this practice. * Justify unidentified chemicals and surrogate standards. FDA expects objective scientific support for how unknown constituents are quantified and why a particular surrogate standard is appropriate. Executive takeaway Toxicological risk assessment should not be treated as an isolated report generated at the end of biocompatibility testing. It is the downstream interpretation of an evidence system that begins with the finished device and depends on how that device is manufactured, extracted, analyzed, quantified, and controlled over its lifecycle. The strongest toxicological conclusion cannot compensate for weak underlying data. Keywords Biocompatibility, toxicological risk assessment, chemical characterization, ISO 10993-18, biological safety, analytical evaluation threshold, AET, extraction replicates, CRO oversight, supplier change control 🎧Listen to the Executive Brief for a closer look at FDA’s expectations for toxicological data, and why extraction design, analytical sensitivity, device variability, and laboratory controls can determine whether a biological safety assessment is truly defensible. Thanks for reading Let's Talk Risk!. If you liked this post, share with others. Note: The audio summary was prepared using Google NotebookLM, an AI-enabled research tool. Here are a few key resources used for this analysis: * FDA (2026). Biocompatibility Town Hall #2: Extractions for Conducting Chemical Analysis for Biocompatibility Assessment. U.S. Food and Drug Administration, Center for Devices and Radiological Health (CDRH). This is a public episode. If you'd like to discuss this with other subscribers or get access to bonus episodes, visit naveenagarwalphd.substack.com/subscribe
Audio and video play straight from the show's own hosting. PodSubs stores and serves no media.
Transcript
From the show. Tap a time to play from there.
SPEAKER_01If you are evaluating a piece of commercial real estate to build a massive skyscraper, the absolute last thing you want to do is pour the concrete foundation without testing the chemical composition of the soil first.
SPEAKER_00Right. Obviously.
SPEAKER_01Because you could have the most brilliant structural engineering on paper, but if there's an undocumented acidic compound hidden deep in that soil, it's just going to quietly eat away at the concrete over time.
SPEAKER_00Yeah, and the entire downstream investment is just structurally compromised before the first steel beam even goes up.
SPEAKER_01Exactly. And you can't salvage a sinking building just by adjusting the top floors.
SPEAKER_00No, absolutely not.
SPEAKER_01The fundamental data supporting that environment, it has to be objective and, well, exhaustive from day one.
SPEAKER_00Which brings us to our listeners today. Because if you are managing a medical device portfolio, or if you're steering a regulatory strategy right now, you are essentially building that skyscraper. And your soil test, the absolute foundational data set, is chemical characterization.
SPEAKER_01That's a really great way to look at it. Because if that underlying extraction data is murky, or if, say, a contract laboratory cuts a corner, your entire biological risk assessment is sitting on very shaky ground.
SPEAKER_00Exactly.
SPEAKER_01So today, we are doing a deep dive into the FDA's recent biocompatibility town hall from September 2026. The agency really shared a wealth of data on how they are evaluating chemical analyses and extraction protocols right now, specifically looking at ISO 10993-18.
SPEAKER_00And the focus of that data, along with the feedback surrounding their recent draft guidance, it really highlights a shift in regulatory expectations.
SPEAKER_01A pretty significant shift, right?
SPEAKER_00Very significant. The agency is heavily scrutinizing the actual physical mechanics of how sponsors are extracting chemicals from these devices. They're looking at solvents, they're looking at temperatures, and really importantly, how variability is calculated.
SPEAKER_01Yeah.
SPEAKER_00It's an incredibly technical area, but it directly impacts whether a submission moves forward or just stalls out completely.
SPEAKER_01Which is exactly why we need to translate those technical laboratory mechanics into executive strategy today. Because frankly, understanding how a mass spectrometer reads an extractable profile isn't just trivia for chemists. I mean, it dictates your speed to market.
SPEAKER_00Right.
SPEAKER_01It dictates how you govern your testing partners and how you build out your ISO 14971 risk management files.
SPEAKER_00Yeah, it really does touch everything.
SPEAKER_01So let's just establish the baseline here. Chemical characterization, or CC, is essentially the gateway to biological safety. It feeds directly into the toxicological risk assessment.
SPEAKER_00The TRA, exactly.
SPEAKER_01Right. We are basically trying to figure out what chemicals might leach out of a device and into a patient.
SPEAKER_00Yeah, that is the core mechanism of the biological risk assessment tree. The FDA's expectation is that you identify and quantify every constituent chemical that could result in human exposure.
SPEAKER_01Every single one.
SPEAKER_00Every single one. But the mechanical difficulty there is that we aren't just looking at the bill of materials, right? We're not just looking at intentional additives or surface coatings.
SPEAKER_01Right. Because things change during production.
SPEAKER_00Exactly. We have to account for degradation products and manufacturing residuals. So that means identifying the trace amounts of cleaning detergents the machining lubricants or even the sterilization gases that sort of hitched a ride during the manufacturing process.
SPEAKER_01And see, from a business perspective, that requirement can quickly sound like a completely open-ended, infinite research project.
SPEAKER_00Oh, absolutely.
SPEAKER_01I mean, if you have to identify every single molecule that touched a device across a global supply chain, you're going to burn through capital and development timelines incredibly fast.
SPEAKER_00Yeah, you would never finish.
SPEAKER_01Right. So the regulatory framework has to have a boundary, which really brings us to how ISO 1099312 standardizes this risk approach.
SPEAKER_00Right. And that boundary is established by focusing on the final finished device. The standard framework is designed around a really fundamental reality, which is that manufacturing changes chemistry.
SPEAKER_01It alters it completely.
SPEAKER_00Yes. So you must account for all materials after they have gone through every manufacturing step, every sterilization cycle, and every anticipated user manipulation.
SPEAKER_01And I really want to break down why that's so critical, because I see this as a major friction point for development teams all the time. Say a supplier sells you a pristine, medical-grade polymer resin.
SPEAKER_00Okay.
SPEAKER_01The second you heat that polymer to mold it into, like, a catheter, you are altering its physical chemical characteristics.
SPEAKER_00Yes, the heat history.
SPEAKER_01Exactly, the heat history. It might cause the polymer to break down into smaller monomer units that could actually leach out. If you just test the raw resin in a vacuum, you completely miss the actual chemical profile of the device you are putting into a patient.
SPEAKER_00And that heat history combined with the chemical interactions of sterilization, that creates what we call the worst case profile. Right. The standard expectation is that you must generate the maximum potential extractables for your assessment. To do that mechanically, you have to utilize the device model or configuration that possesses the largest surface area of each material.
SPEAKER_01Because more surface area exposed to a solvent equals a higher concentration of chemicals pulled into the solution.
SPEAKER_00Exactly.
SPEAKER_01But, okay, here is the strategic bottleneck for executives. Manufacturing hundreds of final finished devices with the absolute largest surface area purely to destructively test them in a lab is incredibly expensive. It's a massive drain on resources really early in the development cycle. So development teams, you know, they often turn to testing coupons.
SPEAKER_00Right. Surrogate coupons.
SPEAKER_01Yeah, which are essentially just surrogate pieces of the raw material. But from everything the FDA is signaling right now, relying on a coupon can be a massive regulatory liability if it isn't documented perfectly.
SPEAKER_00A surrogate coupon is truly only as strong as the gap analysis supporting it. If a sponsor decides to use a coupon instead of the final finished device, they have to embed a rigorous justification within their risk management file. That gap analysis has to prove objectively that the manufacturing differences between the final device and the flat surrogate coupon won't introduce different chemicals or alter the extraction profile.
SPEAKER_01You essentially have to prove a negative.
SPEAKER_00Basically, yeah.
SPEAKER_01You have to demonstrate that the flat coupon went through the exact same heat cycling, the same physical stresses, and the same sterilization as the complex final device. And if that documentation is thin, the FDA is just going to reject the data.
SPEAKER_00And we are seeing the results of those thin justifications in real time.
SPEAKER_01Yeah, the audit data.
SPEAKER_00Exactly. At the town hall, the FDA shared data from their fiscal year 2025 pre-market deficiency audits. They sampled a set of audited letters to give the industry a really transparent look at where submissions are falling short.
SPEAKER_01And what were the numbers on that?
SPEAKER_00Out of a sampled set, 19 audited letters contained 32 specific deficiencies related strictly to chemical characterization and toxicological risk assessment.
SPEAKER_01Wow. I mean, 32 deficiencies across 19 letters. That is a dense concentration of focus. That's almost two severe chemical characterization deficiencies per letter.
SPEAKER_00It is.
SPEAKER_01The FDA doesn't just share that kind of data unless they are actively trying to signal a systemic industry misunderstanding. So let's unpack the specific themes they highlighted there.
SPEAKER_00Well, the top deficiencies were inadequate or unclear extraction conditions.
SPEAKER_01Meaning what exactly?
SPEAKER_00Meaning the lab just didn't use the right solvents or the right temperatures to pull the chemicals out in the first place. They also flagged incomplete test reports and a really high frequency of unidentified chemicals. But perhaps the most complex area of deficiency they noted involved the calculations for critical thresholds. Specifically, the FDA is seeing major issues with how sponsors calculate the analytical evaluation threshold. Okay,
SPEAKER_01let's actually stop and define those thresholds because I think if you're an executive reviewing a multi-hundred page testing report from a contract research organization, a CRO, you really need to know what those terms actually mean in practice.
SPEAKER_00That's fair.
SPEAKER_01Let's start with the analytical evaluation threshold, the AET. As I understand it, the AET is essentially the line in the sand, right?
SPEAKER_00Yes.
SPEAKER_01It's a calculated concentration limit based on toxicological safety. If a chemical extracts at a level below the AET, it's considered too low to pose a toxicological risk and we don't have to spend months trying to identify it.
SPEAKER_00That is the functional application of the AET, exactly. It prevents that infinite research project we talked about earlier by establishing a safe floor.
SPEAKER_01Right.
SPEAKER_00But calculating that floor is highly complex. It involves taking the dose-based threshold, which is the absolute maximum safe dose of a theoretical chemical for a patient, and converting it into an extraction concentration.
SPEAKER_01Based on the device itself?
SPEAKER_00Yeah, based on the medical device's surface area, the volume of the clinical extract, and the number of devices a patient might be exposed to simultaneously.
SPEAKER_01It's a massive math equation that essentially dictates your entire testing burden.
SPEAKER_00It really is.
SPEAKER_01But then you also mentioned the limit of quantification, the LOQ, and the uncertainty factor. How did those interact with the AET? Because if a chemical comes in below the AET, we're supposed to be safe, right? But how do we actually know the lab equipment didn't just miss a dangerous chemical or misread the amount?
SPEAKER_00Well, that is exactly why the uncertainty factor exists.
SPEAKER_01Okay.
SPEAKER_00Right. Right. Right. Right.
SPEAKER_01So you actively lower the AET using the uncertainty factor just to be conservative.
SPEAKER_00Exactly.
SPEAKER_01And then the limit of quantification, the LOQ, that's just the absolute lowest concentration the machine can actually measure with any kind of accuracy.
SPEAKER_00Right. The floor of the machine itself.
SPEAKER_01So if your calculated AET is lower than your machine's LOQ, your machine isn't sensitive enough to prove the device is actually safe.
SPEAKER_00Precisely. If your safety threshold requires you to detect chemicals at, say, one part per billion, but your lab's instrument bottoms out at 10 parts per billion.
SPEAKER_01You're flying blind.
SPEAKER_00Right. Your test is fundamentally inadequate. And the FDA is seeing submissions where these mathematical relationships are just misaligned or where the uncertainty factor wasn't justified by the analytical data at all.
SPEAKER_01Another area they flagged in that audit data was the use of surrogate standards. I'm going to dive into that for a second because it highlights the reality of dealing with unknown chemistry.
SPEAKER_00Oh, absolutely.
SPEAKER_01When a lab finds a mystery chemical in the extract, they have to figure out how much of it is there, but they don't have a pure sample of the mystery chemical to calibrate the machine against. So they have to pick a known chemical, a surrogate standard, that they hope behaves similarly in the machine, and they use that to estimate the concentration of the unknown.
SPEAKER_00Yeah, it's a very sophisticated estimation process. The deficiency arises when the laboratory selects an inappropriate surrogate standard.
SPEAKER_01Like what?
SPEAKER_00Well, if you're trying to quantify a complex, highly polar degradant, but you calibrate your instrument using a simple nonpolar surrogate standard, your estimated concentration will be wildly inaccurate.
SPEAKER_01Oh, wow.
SPEAKER_00Yeah. So the FDA is demanding objective justification for exactly why a specific surrogate was chosen to represent an unknown chemical.
SPEAKER_01You know, if we pull all of these deficiencies together, the delicate threshold math, the surrogate standards, the unidentified chemicals, to me this sends a massive signal for post-market change control.
SPEAKER_00Oh, for sure.
SPEAKER_01If the AET is that sensitive, it means our post-market triggers need to be airtight. A supply chain manager cannot just swap out a manufacturing lubricant to, say, five cents a unit without running it through a biocompatibility lens first.
SPEAKER_00Definitely not.
SPEAKER_01The chemical profile is deeply tied to the continuity of your manufacturing and supply chain processes. A single, undocumented change on the factory floor can literally invalidate years
SPEAKER_00of toxicological safety data. Which naturally leads us to how organizations actually execute these extractions in the first place. Because most device manufacturers do not maintain a fleet of mass spectrometers in-house.
SPEAKER_01No, they contract this work out to specialized testing laboratories or CROs.
SPEAKER_00Right.
SPEAKER_01And this is exactly where QA and RA executives need a really concrete action plan for governing those testing partners. You can't just sign a statement of work, hand over the devices, and assume the resulting report is going to meet current FDA expectations.
SPEAKER_00You really can't.
SPEAKER_01You have to audit their mechanical protocols, particularly around the concept of replicates, which I know the FDA spent a significant amount of time addressing at the town hall.
SPEAKER_00They did. The mandate here revolves around differentiating between an analytical replicate and an extraction replicate. They serve entirely different diagnostic purposes.
SPEAKER_01Break that down for us.
SPEAKER_00So an analytical replicate assesses the variability of the laboratory instrument itself. The lab takes one single device, prepares one batch of extract solution, and then injects that exact same solution into the mass spectrometer three separate times.
SPEAKER_01Okay, so that gives you three different data points, but it only tells you how consistent the machine is.
SPEAKER_00Exactly.
SPEAKER_01It tells you absolutely nothing about whether your manufacturing line is consistently producing safe devices.
SPEAKER_00Which is why the FDA requires extraction replicates to assess device variability. To do this, the lab must take three distinct final finished devices.
SPEAKER_01Three separate devices.
SPEAKER_00Right. And they must subject each of those three devices to its own separate extraction process in its own separate vessel of solvent. This yields three completely independent extract solutions.
SPEAKER_01Okay.
SPEAKER_00The agency's stated best practice expectation To capture robust data while remaining least burdensome is to perform the extraction in triplicate and then analyze each of those three extracts individually.
SPEAKER_01Individually. So three separate devices, three separate jars of solvent, three separate uncompromised instrument readings.
SPEAKER_00Yes.
SPEAKER_01But here is the critical governance issue that we see. Labs are often really tempted to pool those extract solutions before running them through the instrument.
SPEAKER_00Oh, all the time.
SPEAKER_01Right. They take the liquid from jar 1, jar 2, and jar 3, they mix it all together into a single master batch, and then run that single batch through the machine. It saves the lab time, and it saves money on instrument hours. But the FDA explicitly warns against this practice.
SPEAKER_00Because pooling completely masks the device-to-device variability that the toxicological risk assessment relies upon.
SPEAKER_01Think of pooling extracts like taking three separate regional sales budgets. Let's say North America, Europe, and Asia.
SPEAKER_00Okay.
SPEAKER_01You average all three budgets together into one global number. And then you try to look at that one single number to spot a specific localized overspend in Germany.
SPEAKER_00You can't do it.
SPEAKER_01You can't. You have completely erased the variance data. If one out of your three medical devices had a massive spike in a toxic residual chemical, pooling the liquid just dilutes that spike across the clean solvent from the other two devices.
SPEAKER_00The localized overspend simply disappears into the average. And by the time the data reaches the toxicologist, they are looking at an artificially lowered concentration. They might clear a device that actually possesses a dangerous intermittent manufacturing defect.
SPEAKER_01So the executive action plan here is extremely straightforward. Your quality agreements and your statements of work with CROs must explicitly forbid the pooling of extraction solutions.
SPEAKER_00Absolutely.
SPEAKER_01You are paying for variance data to protect your submission and your patients. Do not let a laboratory wash it away for the sake of convenience.
SPEAKER_00And, you know, it is worth noting that for highly novel materials or devices with complex geometries where standard triplicate extraction protocols might not seamlessly apply, sponsors have a strategic avenue available.
SPEAKER_01The QSOB.
SPEAKER_00Yeah. Utilizing the pre-submission or Q-sub pathway allows you to engage the FDA early to lock in your extraction methodology. Taking the time to align on your solvents, your temperatures, and your replicate strategy before you execute the final testing, that is a profound risk mitigation strategy.
SPEAKER_01Absolutely. Do not wait until the final pre-market review to find out your extraction methodology was misaligned with current expectations. Right. Now, to ground everything we've discussed so far, the FDA actually provided two highly specific real-world product exemplars at the town hall. And I think these examples really illustrate how tailored this extraction process has to be to this specific clinical use case.
SPEAKER_00They really do. The first exemplar they analyzed was a blood collection or infusion tubing device. Because a patient might be connected to this tubing for days at a time, it falls under the category of prolonged patient contact.
SPEAKER_01And I noticed they emphasized this specific geometry. Like, if you take a long piece of medical tubing, you can't just chop it up into tiny pieces and throw it in a beaker of solving.
SPEAKER_00No, definitely not.
SPEAKER_01Because if you chop it up, you are exposing the interior layers of the polymer, layers that the patient's blood or fluid would never actually touch in the real world.
SPEAKER_00That is a critical distinction. Chopping the tubing creates an artificial worst-case scenario. It leads to the extraction and identification of chemicals that pose absolutely no real-world clinical risk.
SPEAKER_01Which just unnecessarily clutters the toxicological risk assessment.
SPEAKER_00Exactly. Instead, the mechanically appropriate extraction method is to simulate clinical use. The lab should fill the internal lumen of the tube with the extraction solvents. Typically, this involves using both polar solvents like water or saline and nonpolar solvents to mimic the exact physiological pathway of the blood or fluid.
SPEAKER_01You test the specific surface that actually interacts with the human body. That makes sense. It keeps the data relevant. Now, what was the underlying challenge with the second exemplar?
SPEAKER_00The second exemplar involved a much more delicate technology, a drug-eluting degradable scaffold.
SPEAKER_01Oh, wow. Okay.
SPEAKER_00This represents a long-term contact device, which triggers the requirement for an exhaustive extraction using a full spectrum of polar, semi-polar, and non-polar solvents. But the critical variable here is not just the solvent, it is the temperature of the extraction.
SPEAKER_01Right, because these devices are designed to sit in the human body at 37 degrees Celsius. But labs usually try to speed up extractions by raising the temperature well above 37 degrees. Why is that temperature increase such a massive variable for a degradable scaffold?
SPEAKER_00Well, because a degradable scaffold, by its very nature, is chemically designed to break down over time. Oh, I see.
SPEAKER_01So the extraction must be aggressive enough to capture the true worst case profile, but carefully controlled so it doesn't artificially destroy the material it is trying to characterize.
SPEAKER_00Exactly. It is a massive balancing act.
SPEAKER_01And when you look at the sheer complexity of characterizing an absorbable hydrogel or an in-situ polymerizing material, the executive takeaway points directly back to upstream supplier management.
SPEAKER_00Yes. Supply chain transparency becomes the ultimate business advantage here.
SPEAKER_01Because if you're utilizing materials that already have established device master files, or MAFs, you are building on a known foundation.
SPEAKER_00Right. The supplier has already characterized the material.
SPEAKER_01Exactly. They have already mapped the extraction profile, and the FDA already has that data on file. You drastically lower your biological risk profile. Compare that to the business cost of buying a proprietary, undocumented polymer blend from a supplier who just refuses to share their formulations.
SPEAKER_00Yeah, then you have to reverse engineer and characterize their chemistry from scratch.
SPEAKER_01And the cost difference is just night and day.
SPEAKER_00Which reinforces the necessity of binding quality agreements with your upstream suppliers. You must establish mandatory change notifications for any incoming material controls.
SPEAKER_01Because if they change something.
SPEAKER_00If your supplier quietly changes the resin formulation or even switches their own upstream chemical provider without notifying you, your entire chemical characterization profile is instantly compromised.
SPEAKER_01It always comes back to governance. So let's synthesize the core of what we've unpacked today. Chemical characterization is not just some administrative checkbox you leave to the lab technicians.
SPEAKER_00No, absolutely not.
SPEAKER_01It is a strategic business driver. By integrating worst-case extraction profiling directly into your ISO 14971 risk management files and ensuring your gap analyses for any surrogate coupons are airtight, you build a regulatory architecture that can actually withstand FDA scrutiny.
SPEAKER_00Furthermore, maintaining strict oversight of your contract research organizations, specifically forbidding the pooling of extraction replicates and ensuring they are using appropriate surrogate standards, That guarantees that the data feeding your toxicological risk assessment is both accurate and genuinely representative of your manufacturing variants.
SPEAKER_01And finally, leveraging supplier transparency in master files protects your development timelines and your speed to market.
SPEAKER_00Exactly.
SPEAKER_01The goal of this deep dive was really to give you the strategic executive framing and the technical FDA mechanics to look at your internal processes objectively. You know what thresholds are being scrutinized in the audit data, and you know how to audit a CRO protocol for hidden variables.
SPEAKER_00It's all about proactive governance.
SPEAKER_01It really is. But as we wrap up, I want to leave you with a forward-looking thought regarding all of these physical beaker and solvent extractions.
SPEAKER_00Okay, what's that?
SPEAKER_01Well, the methodology is constantly evolving, right? We are spending massive amounts of capital physically destroying devices in laboratories to see what chemicals leach out.
SPEAKER_00True.
SPEAKER_01But as we look toward the future, what happens when predictive computational modeling and artificial intelligence advance to the point where they can accurately simulate material degradation in silica?
SPEAKER_00Oh, wow.
SPEAKER_01If an algorithm can look at the molecular structure of a drug-eluting scaffold, factor in the heat history of the manufacturing process, and perfectly predict the extractable profile without a single drop of solvent ever being used, will physical triplicate extractions eventually become a relic of the past?
SPEAKER_00The shift from physical chemistry to computational toxicology is certainly on the horizon. It's a fascinating prospect.
SPEAKER_01It is something every regulatory strategist should be keeping an eye on. But until those algorithms are fully validated, make sure your physical foundation inspection is flawless. Thanks for joining us on this deep dive.
Summary “Be guided by your own values and facts. That’s important. That will keep you centered” Quality and regulatory leadership in MedTech is about much more than ensuring compliance. Senior leaders must navigate...
Summary “First you have to understand the technology, what the technology does, what it's capable of, and you have to understand its limitations.” Artificial intelligence is rapidly changing how medical device companies...
Software velocity is not simply a development challenge. It is becoming an enterprise capability. Software is increasingly driving differentiation, recurring revenue, and customer value in MedTech. But many...
Summary “At the end of the day, this business of getting a job, keeping a job, getting promoted is human to human.” The MedTech job market has changed. More candidates are competing for fewer well-matched opportunities,...
PCCPs can turn regulatory cycle time into a competitive advantage, but only if the evidence needed to support future change is funded in advance. Predetermined Change Control Plans (PCCP) are often discussed as a...