Invisible Force: MRI Accident Investigations

S4E1 - “The Probe That Passed the Checks: The Stanford MRI Accident”
A patient goes in for a spine MRI at Stanford Health Care. After the scan, his care team finds a temperature probe melted in his mouth… fused to his lip and tongue. How did it get past the MRI safety checks? In the Season 4 premiere of Invisible Force, John Posh and Tobias Gilk follow the records back to the moments before the scan. The screening form asked about a different form of temperature monitoring. It didn’t ask about the esophageal temperature probe that was there. The probe’s wire was partially concealed in tape, blending into the patient’s breathing tube, and a handheld detector gave no warning. The danger in this MRI accident didn’t depend on large metal objects flying toward the magnet. We explain how radiofrequency (RF) heating can cause a serious burn, even when the thin wire doesn’t move an inch, then ask what it takes to screen a patient who can’t tell the MRI team what’s attached to him. This is the first episode of a three-part investigation into the Stanford MRI burn. Next, we’ll look at what the hospital changed - and how anyone can tell whether those changes worked. Explore episodes and show notes at InvisibleForcePodcast.com. Have information about an MRI accident? Leave a voicemail at 631-674-8477 (631-MRI-TIPS). Follow Invisible Force to continue the investigation. Created and written by: CAIREreporting.org and Tobias “Toby” Gilk Hosted by: Tobias “Toby” Gilk and John Posh Music & Art: Reggie Battle Production: Multitude Website: InvisibleForcePodcast.com