0:00Speaker 1On the third of June nineteen eighty five, a woman entered the Kenistone Regional Oncology Centre in Georgia in the United States. She was there to receive cancer treatment using a state of the art machine, the THERAC twenty five. Unbeknownst to her, however, this machine had two hidden fatal flaws, flows which would cause devastating injury to her and to five other patients before they were finally fixed. Radiation has
0:30Speaker 1been used to treat cancer since the late eighteen hundreds, when it was noted that prolonged exposure to X rays resulted in skin burns. Through a series of trial and error experiments, doctors discovered that it was possible to target this burning effect to remove growths and treat lesions. After much experimentation, medical professionals began to use targeted radiation to destroy cancerous tumors. Modern radiotherapy exposes cancerous cells to radiation, which damages their DNA and destroys these cells, resulting in
1:04Speaker 1the shrinking or complete destruction of tumors. Radiation can also damage the healthy cells which surround a tumor, and so radiation therapy is used in a highly targeted fashion. To maximize damage to cancerous cells and minimize damage to healthy ones. The THEAK twenty five was machine for administering radiation therapy designed and built by Atomic Energy of Canada Limited AECL. It was preceded by the THEAK six and THERRAC twenty,
1:36Speaker 1but was significantly more advanced than either of those machines. Those machines had mostly been operated manually, with technicians physically adjusting the machine for each patient. The THERAK twenty five, by contrast, was fully computer controlled, allowing for much faster treatments to be administered with less human error. The THEAK twenty five had three modes of oper In its field light mode, a mirror moved into position to direct a beam of light, which could be used to ensure that
2:08Speaker 1the patient was positioned correctly. In electron mode, scanning magnets moved into position and spread an electron beam over a wide area, ideal for treating cancers near the surface of the skin. To treat tumors that were deeper within the body, X ray mode would be used. In this mode, a tungsten target was placed in front of the electron beam in such a way that when the beam made contact, it would produce deeply penetrating X ray photons aimed at the patient. On the third of June nineteen eighty five,
2:38Speaker 1sixty one year old Katie Yarborough was due to receive treatment using the THRAC twenty five machine at the Keniston Regional Oncology Centre in Georgia. She had already had a cancerous lump removed successfully from her left breast. This treatment was to irradiate lymph nodes close to the area from which the lump had been removed, in order to prevent her cancer her returning. When treatment began, Katie felt an intense burst of heat, causing her to complain to the machine's operator that she had been burned. At the time,
3:11Speaker 1there was no visible injury where the beam had hit, and so Katie was sent home. Days later, she returned a coin sized burn had appeared. Over the course of weeks, this injury worsened until it was an open wound that resembled a severe radiation burn. She suffered paralysis of her arm and constant pain. Convinced that the THERAK twenty five had caused this injury, Katie began legal proceedings against AECL, who denied any wrongdoing. And settled out of court. Katie
3:43Speaker 1died in a car crash in nineteen ninety. On the twenty sixth of July nineteen eighty five, a forty year old woman with cervical cancer was treated using a THAK twenty five machine to target her hip at the Hamilton Regional Cancer Centre in Ontario. During treatment, she experienced a burning sensation, which was followed by the slow development of redness, swelling, and a wound that resembled a radiation burn. She was admitted to hospital to treat this injury, and it was found that the damage to her hip was so severe
4:15Speaker 1that it would have to be completely replaced. The woman passed away from her cancer before this could happen. AECL sent technicians to inspect the machine at the Hamilton Regional Cancer Centre and concluded that there was a problem with a turntable that placed one of three devices in front of the beam, the mirror for field light mode, the scanning magnets for electron mode, and the tungsten target for X ray mode. If the turntable had been in the wrong position when the beam fired, this would explain the
4:46Speaker 1massive overdose that the patient had received. A fix was made with hardware and software modifications made to the eleven THRAC twenty five machines across the US and Canada. Aecl confidently announced that the MOS machine was now fixed and safer than it ever had been before. Despite this fix, further accidents followed. In December nineteen eighty five, a woman was burned by the THERAK twenty five machine at a clinic in Yakima, Washington, and needed multiple skin grafts to
5:17Speaker 1close the wound. Two further accidents took place at the East Texas Cancer Centre in Tyler, Texas. On the twenty first of March nineteen eighty six, a male patient experienced such a strong and painful burning sensation during treatment with the THEREK twenty five machine that he removed himself from the machine and hammered on the door of the room, demanding to be let out. He was hospitalized shortly afterwards and died less than six months later. On the eleventh of April nineteen eighty six, another male patient was receiving
5:50Speaker 1treatment for skin cancer on his ear. He experienced a sudden, intense burning pain in his head and heard a sizzling sound. Over the course of the next month, his condition deteriorated, and he died from radiation burns to his brain. Following this fatal accident, AECL began an investigation, with their findings reported to the Food and Drug Administration FDA. They made little progress beyond the already identified turntable issue. It was
6:21Speaker 1only when East Texas Cancer Centre physicist Fritz Harger started running tests of his own that the issues underpinning the accidents were brought to light. Harger spoke to the technician who had been using the machine when the accidents occurred and noted that each one had been preceded by the same error message, error fifty four. Harger started trying to recreate that error rather than engaging in slow, careful, systematic testing. He'd used the machine as it might be used on a day to day basis, in putting multiple prescriptions, making
6:54Speaker 1and correcting errors, and importantly doing so as quickly as the average operator. He discovered that should the operator select X ray mode, then quickly go back and switch to electron mode, the turntable would not correctly move into position, prompting an error fifty four message. A further complicating factor was that the THARAC twenty five often produced error messages. These were almost always given as simple numbers with no explanation error one, error two, error three, and so on.
7:30Speaker 1Since most errors were minor and not safety critical, operators were in the habit of clearing error messages and continuing with treatment whenever they occurred. Error fifty four, though, was not trivial. When it occurred, and when treatment continued anyway, the machine could produce an electron beam with the turntable in the wrong position, risking a massive radiation overdose to the patient. This kind of catastrophic error had been impossible with the older THEAK machines. The THERAK six and THEAK
8:02Speaker 1twenty had relied on hardware interlocks, physical mechanisms that prevented the machine from functioning if a dangerous situation existed, for example, if the turntable wasn't properly in place. With the THEAK twenty five, these hardware interlocks were replaced with software based safety checks, which tragically failed. It appeared that AECL had been over confident in their software. They had dedicated a lot of time and energy to testing the physical parts
8:33Speaker 1of the machine, but very little to ensuring that the software worked as it should. Indeed, the code was written in house by a single programmer and was not reviewed by anyone outside the company before it was put into operation. It appeared almost as though AECL had not believed that software could malfunction. Following these discoveries, the FDA declared the THEAK twenty five defective and ordered that it be taken out of operation and d till a fix had been implemented. AECL worked to install software patches on all existing machines
9:06Speaker 1while dealing with multiple lawsuits from the families of those injured. These lawsuits were settled out of court, and the machines returned to service in nineteen eighty six. One error had been fixed, but in January nineteen eighty seven, another different issue came to light. A counter in the machine's code kept track of errors during setup. If the counter contained any number other than zero, there was an error and treatment could not start. However, this counter had an upper
9:38Speaker 1limit two hundred and fifty five, the largest number that could be stored in a single byte of data. If the counter reached a value above two hundred and fifty five, it would reset to zero, allowing treatment to begin even if there were still errors. In January nineteen eighty seven. This was exactly what happened a THRREK twenty five five malfunctioned again, resulting in another patient death. This final accident was the end for the THEAK twenty five. It was considered
10:10Speaker 1by most institutions to be too dangerous to carry on using, and the majority of machines were soon retired. Modern radiotherapy machines do not have the same flaws as the THEAK twenty five. They don't rely solely on software and instead have mechanical interlocks that physically stop the machine from activating if a dangerous condition exists. All software for medical devices is subject to external review and must conform to pre agreed standards. The THEAK twenty five, a machine that was
10:43Speaker 1meant to heal, but which in a small number of cases did the exact opposite, is now a case study for engineers around the world. It demonstrates the importance of mechanical safety features and the dangers of putting too much faith in software.