Company Spotlight: Crescom MediAI-OA - Automating Knee Osteoarthritis X-Ray Assessments
We unpack the FDA 510(k) summary for Crescom's MediAI-OA, an AI tool designed to automate Kellgren-Lawrence grading and structural assessments for knee osteoarthritis. We explore the performance data, the demographics of its test set, and what health-system buyers need to know before deploying it.
Key points
MediAI-OA is an automated image processing software cleared to aid in assessing radiographic knee osteoarthritis using Kellgren-Lawrence grading and OARSI criteria.
Clearance was based on substantial equivalence to a predicate device, Alpha Intelligence Manifolds' DeepXray.
In an independent test set of 1,173 subjects, the software achieved a sensitivity of 0.876 and a specificity of 0.865 for detecting a Kellgren-Lawrence grade of 2 or higher.
The independent test set was heavily skewed demographically, consisting of 83.8% White subjects and only 1.2% Asian subjects.
Testing relied almost entirely on computed radiography (CR) images rather than directly digital (DX) images, and several major X-ray hardware brands present in the training set were completely absent from the independent test set.
Source: FDA 510(k) summary K253172: MediAI-OA (Crescom) - U.S. Food and Drug Administration, 2026
This episode is an AI-generated conversation summarising a public document; the hosts' voices are synthetic. It is for information only and is not medical advice. Always refer to the original source.
Company: Crescom · Product: MediAI-OA · FDA 510(k) K253172
Full transcript: https://ai-in-medicine-podcast.vercel.app/episodes/company-spotlight-crescom-mediai-oa-automating-knee-osteoarthritis-x-r-50290c